Toho-electronics TRM-006A Manual de usuario Pagina 24

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 40
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 23
24
g) Data portion
This specifies data to be written in the register. The data is fixed at 4 bytes. This product
will write data without the decimal point.
Example: In the case of numerical data
Example Significance of the value
Proportional band (P) = 1.0 % 0000000AH
PV = 1200.0°C 00002EE0H
SV = -10.00°C FFFFFC18H
In the case of text data, write the ASCII code "INP" ( is a space): 20494E50H.
h) Error code
If a message from a high-level computer is error-ridden, it will be incorporated in the
"response message" from this product and returned.
The error number "04" is an instrument error (memory error or A/D conversion error, AT
error). It will be incorporated in the "response message" regardless of whether there is an
error in the "request message."
If there are two or more errors occurring at the same time, the largest error number will be
incorporated.
The table below indicates the error contents and classifications.
Error No. Error contents in the "request message" received by this product
01 Received an unsupported function code.
02 Received an address other than the specified one.
03 The numerical data deviated from the "range of settings designated specifically with
setting items."
04 Instrument error (memory error or A/D conversion error, AT error)